Reports until 18:43, Wednesday 20 December 2023
cSQZ General
jacques - posted 18:43, Wednesday 20 December 2023 - last comment - 17:21, Thursday 21 December 2023(11451)
PLL debugging (1): strange PZT driver

Yesterday with Eric we noticed a very peculiar phenomenon happening when the correction sends a DC offset (so that the PZT is at the middle of its range), then the PID is switched on, then off, while the DC offset is kept on. Figure 1 shows the error signal when only the DC is sent. This is strange because there are stable, periodic squarish oscillations at ~6 kHz, while no actuation is being done. This phenomenon appears whether the PID is being done by the RP or by a Newfocus LB1005 analog servo controller, which means that it could not have come from the PID block in itself. Furthermore it appears whether I use channel X or channel Y of the PZT driver

Today I put the correction PID back to the RP and compared, for fixed {offset = 0.5, P = 1e-4, I = 1, D = 0}, the PLL's Thorlabs PZT driver with the one used to lock the csqz. Figures 2 and 3 show the time domain error signal and the frequency domain PD signal for the csqz PZT driver, while figures 4 and 5 show them for the PZT driver used during the previous days.

Conclusion: the weird 6kHz oscillation comes from the PZT driver, though the exact phenomenon by which it appears remains unknown. We should use another PZT driver.

Images attached to this report
Comments related to this report
jacques - 17:21, Thursday 21 December 2023 (11456)

Hypotheses which can be tested in the future to improve the loop:

1/ On the weird sidebands which appear on the PD signal when the loop is closed with the RP: it is possible that these come from the VCO. A test could be to swap the RF generator back to the Moglabs at the same 52MHz frequency and check again for the presence of these sidebands.

2/ On the stability of the loop: one possible cause of instability is that the proportional gain is too high. I cannot reduce it further because of internal RP contraints. Furthermore the noise figure of the 10x amplifier is unknown; it is possible that this components adds too much noise to the correction. A test could be to swap out the 10x amplifier. The drawback is that we will only have ~10V of actuation range vs 75V initially but this can be managable for the short term, where we are only interested in having a stable loop for a short while.