I closed the loop of the PLL for the first time by coding a PID on the RP (with P = D = 0, I = 10) and sending the output to the 10x amplifier then into the PZT driver. The output voltage is between 4 and 65V to make sure that the PZT port of the DBM is not damaged.
I preemptively made sure that the beatnote of ~55MHz can be seen around the midvoltage (~34V) in order to maximize the range of the PZT drive.
Figure 0 shows the spectrum of the PD beatnote when the loop is not closed.
Figure 1 shows the spectrum of the PD beatnote signal (at ~55MHz) once the loop is closed and figure 2 shows an example of the PFD "ERR OUT" port.
To test the robustness of the loop I slightly changed the temperature of the lightwave laser by turning its temperature knob and I could see on the spectrum analyser that the plateau at 55MHz stayed in place.
For reference, figure 3 is a spectrum of "ERR OUT" after being attenuated by 6+5+9=20dB
The next step should be to:
1/ better understand these figures and improve the loop's response.
2/ send the error signal's low frequency to a PID then to the temperature control of the DBL