posted 11:19, Monday 18 December 2023 (11446)
Tuning the cavity for second dataset
- Tuned cavity+steering mirrors to try to change cavity splitting to take second dataset
- Initially tried to move mirror C + steering mirrors, but that only shifts the splitting by small amounts (up to 180kHz). Moved different combinations of mirrors to achieve best splitting change.
- Unscrewed and slowly moved out mirror B, and then tuned mostly mirror A (and C when needed) to compensate and close cavity path. Optimized steering mirrors and modematching lenses to achieve maximal possible coupling into cavity. Resulted in a change of 1MHz in cavity splitting ⇒ splitting moved from 12.1 to 13.2 MHz
- Note: Hit a bad spot on either mirror B/D that resulted in some loss? Saw very low cavity transmission. Slowly and iteratively changing the cavity spot resulted in normal transmission again.