swadha - posted 11:19, Monday 18 December 2023 (11446)
Tuning the cavity for second dataset
Tuned cavity+steering mirrors to try to change cavity splitting to take second dataset
Initially tried to move mirror C + steering mirrors, but that only shifts the splitting by small amounts (up to 180kHz). Moved different combinations of mirrors to achieve best splitting change.
Unscrewed and slowly moved out mirror B, and then tuned mostly mirror A (and C when needed) to compensate and close cavity path. Optimized steering mirrors and modematching lenses to achieve maximal possible coupling into cavity. Resulted in a change of 1MHz in cavity splitting ⇒ splitting moved from 12.1 to 13.2 MHz
Note: Hit a bad spot on either mirror B/D that resulted in some loss? Saw very low cavity transmission. Slowly and iteratively changing the cavity spot resulted in normal transmission again.