Reports until 18:29, Friday 15 December 2023
cSQZ General
jacques - posted 18:29, Friday 15 December 2023 (11444)
PLL RP standalone test

I have tested the RP which will be used for the PLL. It is currently misleadingly named shg.mit.edu.

The RP outpout channel 1 is a signal bounded by 0 and 1V. 

I tested a voltage scan at output1, which was fed to a 10x amplifier, then to the Thorlabs PZT driver. The gain and offset have been adjusted so that the voltage out of the driver (which will be ultimately fed to the DBL) is always between 5V and 80V. We are well inside the PZT drive range of the laser (-100V to 100V).

Note for the future: the bandwidth of the 10x amplifier seems to be around 5-10kHz so any correction signal with a frequency higher than this will be filtered.

For amplitude/frequency reference: the signal of the figure is the output of the 10x amplifier, whose input is the RP output channel 1, with the commands  

scopeDecimation =  4096
    # Scan Cavity Mirror
    scan_pll(scanFreq =  1*1/(8E-9*(2**14)*scopeDecimation), 
         sweepAmplitude = 0.45, offset = 0.05)

Next steps: add the PID module to the RP script and close the PLL loop.

Images attached to this report