Following the preliminary test of the PFD (Link) I swapped the {minicircuits TVA-R5-13, 35dB amplifier and the -20 dB attenuator} with a Minicircuits ZFL-500-LN+ of design gain 20dB on the signal of the PD. The new electronic scheme is: PD-> RF Splitter (for monitoring the PD signal on an oscilloscope) -> ZFL Amplifier -> "RF In" port on the PFD. See figures 1 and 2 for pictures of the minicircuits components
Figure 3 shows the signals from the RF splitter (PD signal, channel 2 on the oscilloscope) and the ERR OUT of the PFD (channel 1). The beatnote has a period of 7.5 us so a frequency of 267kHz.
Note that the AOM and LO signals were generated by the moglabs channels 1 (+31dBm) and 2 (+10dBm) at 60MHz.
Conclusion: there is enough power to conveniently generate the error signal at the output of the PFD.