I used the moglabs channel 1 to drive the AOM at 60MHz sharp, 31dBm; channel 2 as the input of the "LO in" of the mixer, 60MHz sharp +10dBm. I optimized the light power arriving on the PLL PD (total power 226uW, comparable to the reference in entry Link ). I amplified the PD signal using the +35dB amplifier (figure 1) and added a 20dB attenuator (otherwise the output of the amplifier saturates the PFD).
To get the laser's beatnote (=PD signal) as close as 60MHz as possible, I fine-tuned the temperature of the lightwave laser around 35°C (the Diabolo is kept at 35°C with a -0.59V external temperature voltage control).
Figure 2 shows the obtained signals: channel 1 (yellow) is the output "ERR out" of the PFD. It has a period of 10us so the beatnote frequency f_PD - f_LO = -2*1/10e-6 = -200kHz. To understand the sign convention and the extra factor 2, look at entry Link. Channel 2 is a pickoff from the PD signal, a zero mean sine wave at 60.2MHz.
Conclusion: following entry Link, we are able to obtain the error signal using the actual PD signal.
Next step: test the Intra-Action VCO model DE-40M and swap it with the Moglab channels.