[Eleonora, Eric]
We installed the GPIB on the AG4395A in NW22.
We acquired the noise spectrum of the SR560 set in DC with G = 100 after debagging several coding and setting issues.
Fig. 1 : spectrum of the SR560 acquired putting 50 Ohm terminators to both A and -B channels of the servo. In blue the spectrum acquired with the AG4395A compared with the spectrum acquired with the SR785 as a sanity check.
Fig. 2: TF acquired with the AG94395A. One cannot go lower than 10Hz with this network analyzer. Power of the sweep sine = -50dBm. The TF was acquired with the servo out of loop. There is an internal LPF in the SR560 at hundreds of Hz.
Fig. 3 : Noise of the SR560 derived as: noise = spectrum / TF. Up to 100 Hz I used the noise measured before (see entry 11393). The difference is at high frequency since now the SR560 is used in a DC configuration and not with a LP.