hudsonl - posted 17:31, Tuesday 05 December 2023 - last comment - 11:51, Friday 08 December 2023(11420)
CSQZ Fiber Box LO Fiber
The LO fiber out of the Fiber box looks to be very lossy. There's 180mW into the fiber box, and we expect the LO fiber to have roughly 90% of this power, or 160mW. Its output is only about 35mW, and the fiber is clearly glowing near where it's spliced when viewed with an IR camera.
Comments related to this report
hudsonl - 14:48, Wednesday 06 December 2023 (11422)
The LO power is also about 25% of the expected power given the 180mW input power. The LO power is 2.5mW vs 9mW expected. This indicates that the problem is not with the new splice, but with a fiber element common to both the LO and the SHG pump paths.
hudsonl - 11:07, Friday 08 December 2023 (11427)
It looks like the 99/1 fiber splitter is damaged. Connecting the PM fiber out of the laser to the fiber splitter, we get about 80mW out of the 99% port with 220mW in, or about 36% coupling. Using a standard single mode patch cable, we get 180mW out, or 82% coupling, more in line with what I would expect. Given this, I believe we should try replacing the 99/1 splitter before we order new AOMs for the fiber box.

The fibers have a damage threshold of about 1 MW/cm^2, and a "practical safe level" of about 250 MW/cm^2 according to thorlabs (see https://www.thorlabs.com/newgrouppage9.cfm?objectgroup_id=1362&pn=P1-1064-FC-1). The fiber we're using has a mode field diameter of about 6 microns, or an effective area of about 1.1 x 10^(-6) cm^2, which gives a damage threshold of about 300 mW and a "practical safe level" of about 75 mW. Based on this, it seems likely we damaged the fiber splitter by sending too much power at it.
hudsonl - 11:17, Friday 08 December 2023 (11428)
While the fiber daage threshold is quoted at 300 mW, the spec sheet for the 99/1 fiber beam splitter says it should be able to handle up to 1 Watt of power. See https://www.thorlabs.com/drawings/830d2b7c309f8c74-E4C1FA78-A61A-FC1E-0E705545D0A363D3/TN1064R1A1B-SpecSheet.pdf
hudsonl - 11:51, Friday 08 December 2023 (11429)
Connecting the fiber-coupled output of the laser directly to the input of the fiber box, we only have about 25% transmission from the input to the pump fiber. We send in about 205 mW and get about 55 mW on the pump path. This indicates that while the 99/1 beam splitter may need to be replaced, there is a sperate problem with the fiber box as well.

Connecting the output of the 99/1 to the input of the fiber box, we have about 200mW in, about 100mW out of the "99%" path, and about 20mW out of the pump path fiber from the fiber box, so the losses all add up, and it looks like there is a problem both with the 99/1 beam splitter and with the fiber box.

The 1% pickoff port of the 99/1 has 1mW of power, so it looks like both paths have about 50% loss. This makes sense, as the damaged fiber looks to be the input to the beam splitter.