GRAVITES General
epolini - posted 20:01, Sunday 03 December 2023 - last comment - 18:02, Monday 04 December 2023(11416)
Measurements with servo in DC both with SR785 and 4395A

On Friday, Eric and I realized that the correction signal of FNC loop was flat at higher frequencies (i.e. 1MHz with the Agilent 4395A). We realized that it might be due to the noise of the SR560 which was set to a LP filter @ 3kHz, as shown by the noise spectrum in the entry 11393 . We changed the configuration of the servo to DC, still with a gain = 100. The purpose of this study is to verify if we are limited by GAWBS as we suspected from the latest study we showed in the entry 11396.

I acquired the error signal (closed loop), correction signal and TF of the system, both with the SR785 and 4395A (to be able to verify the calibration).

Fig. 1: Error signal measured with SR785 (blue) from 1Hz - 102 kHz and measured with the Agilent 4395A (orange) from 1Hz - 1.6MHz.

Fig. 2: Correction signal measured with SR785 (blue) from 1Hz - 102 kHz and measured with the Agilent 4395A (orange) from 1Hz - 1.6MHz.

Fig. 3: OLTF of the system measured with SR785 (blue) from 100Hz - 102 kHz and measured with the Agilent 4395A (orange) from 100Hz - 1.6MHz.

Next step:

Images attached to this report
Comments related to this report
epolini - 18:02, Monday 04 December 2023 (11418)

[Eric, Eleonora]

The discrepancy between the TF acquired with the SR785 and Agilent 4395A is due to the fact that the R channel used for the TF (A/R) had a 20dB attenuation. So the discrepancy was 20dB - 6dB that comes from the 50Ohm input impedence. We verified that the two instruments agree.