swadha - posted 12:15, Saturday 02 December 2023 (11415)
Measurement process and first measurement for ADBC

Measurement process:

  1. Cavity:
    1. Turn down laser power (input l/2 306deg)
    2. Switch to p-pol (pre-vacuum chamber l/2 45deg, refl post vacuum chamber l/2 65deg)
    3. Sweep laser PZT with triangle wave and check that the 00 mode is being maximally transmitted.
    4. Note down cavity coupling (for coupling factor into cavity)
  2. Calibration:
    1. Turn laser power back up (input l/2 340deg)
    2. Switch to s-pol (pre-vacuum chamber l/2 90deg, refl post vacuum chamber l/2 110deg)
    3. Measure P_in with power meter, maximize and note down
    4. Change pre-vacuum chamber l/2 to ~110deg and note down theta_in
    5. At transmission, note down max transmission to PDA100A and then change transmission l/2 to send ~1mW power to 1811 for LO. Note down new PDA100A value to calculate e_out, P_LO
    6. Use Agilent to measure transfer function while driving the EOM (using backwards RF splitter to add to PDH sidebands). Note down modulation depth of calibration sideband, V_ag.
    7. Perform calibration analysis and verify that it matches the model
  3. Noise:
    1. Remove splitter at EOM driver input (disconnect from Agilent because it causes noise)
    2. Change theta_in back to pure s-pol input
    3. At transmission, note down max transmission to PDA100A and then change transmission l/2 to send ~2mW power to 1811 for LO. Note down new PDA100A value to calculate e_out, P_LO
    4. Take spectrum of 1811 upto 20MHz and check that it is clean and at shot noise
    5. Put 1811 AC output into scope and check that it is less than 300mVp (HF2LI range)
  4. Measurement:
    1. Put 1811 AC output into HF2LI and connect HF2LI to laptop
    2. Perform measurement for ~3.5 hours with T_fft = 0.28s, f_res = 3.5 Hz
       

First measurement data: calibration, noise, measurement

First measurement preliminary result with SNR = 1 approach for calculating axion-photon coupling bounds: plot