Measurement process:
- Cavity:
- Turn down laser power (input l/2 306deg)
- Switch to p-pol (pre-vacuum chamber l/2 45deg, refl post vacuum chamber l/2 65deg)
- Sweep laser PZT with triangle wave and check that the 00 mode is being maximally transmitted.
- Note down cavity coupling (for coupling factor into cavity)
- Calibration:
- Turn laser power back up (input l/2 340deg)
- Switch to s-pol (pre-vacuum chamber l/2 90deg, refl post vacuum chamber l/2 110deg)
- Measure P_in with power meter, maximize and note down
- Change pre-vacuum chamber l/2 to ~110deg and note down theta_in
- At transmission, note down max transmission to PDA100A and then change transmission l/2 to send ~1mW power to 1811 for LO. Note down new PDA100A value to calculate e_out, P_LO
- Use Agilent to measure transfer function while driving the EOM (using backwards RF splitter to add to PDH sidebands). Note down modulation depth of calibration sideband, V_ag.
- Perform calibration analysis and verify that it matches the model
- Noise:
- Remove splitter at EOM driver input (disconnect from Agilent because it causes noise)
- Change theta_in back to pure s-pol input
- At transmission, note down max transmission to PDA100A and then change transmission l/2 to send ~2mW power to 1811 for LO. Note down new PDA100A value to calculate e_out, P_LO
- Take spectrum of 1811 upto 20MHz and check that it is clean and at shot noise
- Put 1811 AC output into scope and check that it is less than 300mVp (HF2LI range)
- Measurement:
- Put 1811 AC output into HF2LI and connect HF2LI to laptop
- Perform measurement for ~3.5 hours with T_fft = 0.28s, f_res = 3.5 Hz
First measurement data: calibration, noise, measurement
First measurement preliminary result with SNR = 1 approach for calculating axion-photon coupling bounds: plot