Calibration calculations and measurement
- Calibration scheme:
- We lock the laser in s-pol and scan p-pol sidebands.
- At the output, some of the locked s-pol field is converted to p-pol using a HWP. This acts as the LO. HWP angle is such that most of the signal and some of the LO is sent to the readout PD (Newport 1811).
- To scan, we use the network analyser (Agilent 4395A) as the source that drives the EOM driver circuit ⇒ We use this to scan sideband frequency
- Half wave plate at cavity input is used to send in a superposition of s- and p-pol into the cavity, so that s-pol carrier and p-pol sideband at the appropriate frequency can be simultaneously resonant.
- We measure the transfer function A/R and match it up with our model.
- Input power is 800mW. At each scan, need to measure parameters: theta_in, readout efficiency, P_LO.
- Calculations for model attached.
- epsilon = mode-matching efficiency x signal readout efficiency. Latter to be measured at each scan, former measured using cavity coupling and finesse data, found to be 0.82.
- Measurement vs. model plot attached, seem to match up well.