swadha - posted 15:54, Thursday 09 November 2023 (11405)
Calibration calculations and measurement
Calibration scheme:
We lock the laser in s-pol and scan p-pol sidebands.
At the output, some of the locked s-pol field is converted to p-pol using a HWP. This acts as the LO. HWP angle is such that most of the signal and some of the LO is sent to the readout PD (Newport 1811).
To scan, we use the network analyser (Agilent 4395A) as the source that drives the EOM driver circuit ⇒ We use this to scan sideband frequency
Half wave plate at cavity input is used to send in a superposition of s- and p-pol into the cavity, so that s-pol carrier and p-pol sideband at the appropriate frequency can be simultaneously resonant.
We measure the transfer function A/R and match it up with our model.
Input power is 800mW. At each scan, need to measure parameters: theta_in, readout efficiency, P_LO.
epsilon = mode-matching efficiency x signal readout efficiency. Latter to be measured at each scan, former measured using cavity coupling and finesse data, found to be 0.82.
Measurement vs. model plot attached, seem to match up well.