Finesse measurements and powering up
- Re-measured p-pol finesse for new laser/cavity alignment:
- Drove laser PZT with triangle wave far enough to get two 00 peaks in one ramp. Used distance between two FSRs to calibrate linewidth, PZT transfer function.
- Finesse = 160
- PZT = 5.24 MHz/V
- Re-measured s-pol finesse with cavity storage time method:
- Turned up the cavity input power for first measurements: 70mW -> 800mW