GRAVITES General
epolini - posted 17:46, Tuesday 31 October 2023 (11396)
Simulink model for the updated system (Marconi and SR560)

[Eleonora, Eric]

We adapted the model on Simulink to the new configuration of the system.

PID SR560

Zeros: []

Poles: 2*pi*[3e3]

Gain: 1500000 (*)

VCO Marconi

Zeros: []

Poles: -2*pi*[0]

Gain: 2*pi*[1.5e3]

Previously, with the minicircuits VCO and the LB1005 we set these parameters to:

PID LB1005

Zeros: -2*pi*[3e3]

Poles: -2*pi*[0]

Gain: 0.0174

VCO minicircuits

Zeros: []

Poles: -2*pi*[0]

Gain: 2*pi*[4.028e6]

(*) We needed to set this gain very high, instead of the measured and expected gain = 100, in order to match the simulated OLTF with the measured OLTF. 

We notice that the GAWBS was higher than the correction and error signals measured by a factor 5.

Two observations on this point:

1. Chris is also not confident in the model included in the Optica paper

2. the slope of this noise at high frequency matches perfectly our measurements (scaled by a factor 5).

In Fig. 1 and Fig.2 the error and correction signal plots without adjusting the GAWBS are shown respectively. 

In Fig. 3 and Fig. 4 the  error and correction signal plots dividing the GAWBS by factor of 5 are shown respectively. 

Images attached to this report