[Eleonora, Eric]
We adapted the model on Simulink to the new configuration of the system.
PID SR560
Zeros: []
Poles: 2*pi*[3e3]
Gain: 1500000 (*)
VCO Marconi
Zeros: []
Poles: -2*pi*[0]
Gain: 2*pi*[1.5e3]
Previously, with the minicircuits VCO and the LB1005 we set these parameters to:
PID LB1005
Zeros: -2*pi*[3e3]
Poles: -2*pi*[0]
Gain: 0.0174
VCO minicircuits
Zeros: []
Poles: -2*pi*[0]
Gain: 2*pi*[4.028e6]
(*) We needed to set this gain very high, instead of the measured and expected gain = 100, in order to match the simulated OLTF with the measured OLTF.
We notice that the GAWBS was higher than the correction and error signals measured by a factor 5.
Two observations on this point:
1. Chris is also not confident in the model included in the Optica paper
2. the slope of this noise at high frequency matches perfectly our measurements (scaled by a factor 5).
In Fig. 1 and Fig.2 the error and correction signal plots without adjusting the GAWBS are shown respectively.
In Fig. 3 and Fig. 4 the error and correction signal plots dividing the GAWBS by factor of 5 are shown respectively.