GRAVITES General
epolini - posted 17:43, Monday 30 October 2023 - last comment - 15:13, Sunday 05 January 2025(11391)
Marconi Transfer Function measurement

Today I measured the Marconi's transfer function, which corresponds to the calibration factor from Volts to Hz. I sent a voltage provided by a DC power supply to the external modulation port. The input volage taken by the Marconi accordinag to the datasheet is: 1V RMS (1.414 V peak) sine wave for set deviation and maximum safe input is ±15 V. Fig.1 shows the frequency output of the Marconi measured with the frequency counter as a function of the input voltage. The FM deviation during the measurement was set to 2kHz.

The fit gives a calibration factor of 1.48kHz/V.

Images attached to this report
Comments related to this report
- 15:13, Sunday 05 January 2025 (11853)

Similar measurements on Vienna Marconi 2026B gain vs. modulation deviation span:

Span Gain [Hz/v]
500Hz 710
1kHz 1420
5kHz 7100
50kHz 71100

Haocun