Today I measured the Marconi's transfer function, which corresponds to the calibration factor from Volts to Hz. I sent a voltage provided by a DC power supply to the external modulation port. The input volage taken by the Marconi accordinag to the datasheet is: 1V RMS (1.414 V peak) sine wave for set deviation and maximum safe input is ±15 V. Fig.1 shows the frequency output of the Marconi measured with the frequency counter as a function of the input voltage. The FM deviation during the measurement was set to 2kHz.
The fit gives a calibration factor of 1.48kHz/V.
Similar measurements on Vienna Marconi 2026B gain vs. modulation deviation span:
| Span | Gain [Hz/v] |
| 500Hz | 710 |
| 1kHz | 1420 |
| 5kHz | 7100 |
| 50kHz | 71100 |
Haocun