[Eleonora, Eric]
I summarize here the results of the simulink model interfaced with the measured noises of the system.
Fig. 1: Simulink model of the system / control loop. I underline the noise added at each port (NB. we did not add yet seismic and acoustic noises).
Fig. 2: Error signal closed loop. In black we have the error signal closed loop measured in our system and we see that we are limitedby the PID servo noise.
Fig. 3: Correction signal. In black we have the correction signal we measured in our system. The VCO noise is limiting our system and this is why we cannot suppress more noise. Above 10kHz we might be limited by the PID servo noise.
Since the signal is limited by the servo controller and the actuator (VCO) we can replace these elements with lower noise ones and perform the same measurement in order to understand what would limit us instead. According to the simulation, we should be limited by laser frequency noise.
We are planning to substitute the VCO with a Marconi and the LB1005 with a SR560 and eventually a low pass filter.