I measured the powers circulating in the MZI.
Before 200 meters fiber = 117 uW
After 200 meters fiber = 52 uW
Before 20 cm fiber = 400 mW
After 20 cm fiber = 52 uW (there is a filter in front of the collimator ND 0.2 and I misaligned the collimator to have the same power at the end of both fibers).
I connected the DC channel of the PD 1811 to the oscilloscope and I have around 80 mV per arm, and 160 mV totally in this configuration. I measured with the powermeter the power in the BS output ports: 37 uW in the port going to the PD and 37uW in the other port.
The RF channel pk-pk in this configuration gives 600 mV.
I keep this power level for the next measurements since the error signal is more sinusoidal wrt before. The pk-pk amplitude of the demodulated (after the mixer) signal is 360mV, i.e. the sensor TF is 0.18 V/rad (instead of 0.25 V/rad used before).