Reports until 20:53, Monday 23 October 2023
GRAVITES General
epolini - posted 20:53, Monday 23 October 2023 (11381)
Phase lock VCO to measure VCO noise

Today I measured the VCO phase noise by phase locking it to a RF generator, as Eric suggested. Since we understood that the VCO is the limiting noise at low frequency in the unbalanced MZI system, it was worth it to measure VCO noise properly. To do so we phase lock the VCO and then we know that the correction signal of this loop corresponds to the VCO noise (at least below few kHz) since it is the noisier component of the system. I connected the LO from the RF generator and the VCO output to a mixer. After the mixer the signal is low passed with a 1.9MHz LP filter. The signal then goes to the LB1005 (we wanted to use the SR560 but we need to send a DC voltage to the VCO in order to set its frequency to 71MHz) and then inputed in the VCO. The LB1005 settings are: PI corne 3KHz, LF gain limit 70dB. I locked the servo and then acquired the spectrum with the SR785 (up to 50kHz because then there was a random problem saving data). I calibrated the spectrum from Vrms/rtHz to rad/rtHz using the V -> Hz calibration measured in the entry 11380 and then Hz -> rad diving by the frequency vector. The VCO noise and the correction signal of the system are shown in Fig. 1. There is still some work to do to match the model and the system but we are getting closer. 

Images attached to this report