GRAVITES General
epolini - posted 20:11, Monday 23 October 2023 (11380)
VCO transfer function measurement

Today I measured the VCO TF (calibration), to investigate possible discrepancies between the Smimulink simulation and the system. Before I estimated this parameter from the plot Frequency vs Voltage on the VCO datasheet (minicircuits ZOS 100+). I sent different known voltage values to the VCO using the TENMA DC power supply and I read the output frequencies with a frequency counter (53131A ) borrowed from a shelf in green lab. The plot is shown in Fig. 1 and we get  Frequency_out [MHz] / Voltage_in [V] = 4.028 MHz / V (while before I was using 4.2 MHz / V, which is not so far). 

Images attached to this report
Displaying report 1-1 of 1.