Servo Controller noise measurement (PID LB1005)

Today, I measured the noise of the LB1005 servo controller, to include it in the Similink noise budget. I summarize here the steps, discussed with Eric. 

1.  Measure the PID spectrum 

I disconnected the PID from the loop and connected 50 Ohm terminations to both the PID inputs (A and -B). I connected the PID Out to the oscilloscope and tuned it to zero with the input signal set knob. Then I connected it to the spectrum analyzer (SR785) and I acquired the spectrum for the following settings: PI corner = 3kHz, LF gain limit = 60dB and Output set = 7.8V (needed to drive the VCO). Fig. 1 shows the spectrum calibrated in rad/rtHz ([V]/0.25). 

2. PID TF - out of loop

I measured the TF of the servo controller sending the swept sine to channel A on the LB1005 (100Hz - 100kHz). The SR785 source out was also sent to the channel A of the SR785 as input point of the TF. The OUT of the servo was sent to the channel B of the SR785. During this measurement, the servo controller was set to LFGL, i.e. the P-I filter output is enabled, but the low frequency gain is limited according to the LF Gain Limit switch. (This mode is useful for acquiring lock and situations where limiting the DC gain is necessary), with same settings as during the spectrum measurement. Fig. 2 shows the magnitude and phase of this TF. (NB. the TF needs to be set in dB and not dB/rtHz. To remove the PSD units one needs to go to FFT mode and remove them and then back to swept sine mode).

3. Noise PID Servo Controller 

The nosie of the servo is finally derived by: Noise PID = Spectrum PID / TF. 

Since the TF is from 100Hz I needed to interpolate the data and then linearize the TF: TF_lin = 10^(TF_dB/20). Fig.3 shows the noise asd of the PID in rad/rtHz.

Images attached to this report
Comments related to this report
epolini - 16:22, Sunday 22 October 2023 (11378)

I compared the PID TF measured here with the PID TF derived from the open loop transfer function of the system (OLTF). 

I know that the system is composed by Plant (MZI), Sensor (PD), Actuator (VCO), Filter (PID Servo). The OLTF = P(s) * S(s) * A(s)  * F(s), therefore F(s) = OLTF / ( P(s) * S(s) * A(s) ).

From the output of the PD I can derive P(s) * S(s): V_out = A * sin( x[rad]) ~ A * x[rad], meaning that A is the TF P(s) * S(s) is V_out / x[out] = A, where A is the amplitude of the signal read by the oscilloscope.

A(s) is the TF of the VCO, whose output is in frequency and the input is in Volt : [Hz] / [V] = 4.2e6 Hz / V according to the datasheet. The calibration in radians is obtained diving by the frequency.

The PID TF is obtained as:

TF_PS_VCO_dB = 20*np.log10(TF_PS*TF_VCO)
TF_PID = OLTF_dB - TF_PS_VCO_dB

The resulting PID TF is shown in Fig.1, where the blue curve consider a OLTF with a LF gain limit = 90dB and UGF = 16.8kHz and the orange curve LF gain limit = 60dB and UGF = 18.6kHz (i.e. same configuration as the PID TF measurement of this week). The curve does not match exactly the measurement but maybe there is a small difference in the gain of the PID during the measurements. We wanted to verify that the low gain of the measured PID TF made sense and we can conclude that it does.