- posted 18:43, Wednesday 18 October 2023 (11373)
Measuring the optical path length of Newport EOM for 532
Masaya

To design the layout of optics towards PDH lock of SHG of cSQZ,we needed to figure out the optical path length. I measured beam size in the cases with and without EOM. As a result, the ray transfer matrix(ABCD matrix) of the EOM is

{{1,2.3cm},{0,1}}for x-axis, {{1,3.9cm},{0,1}}for y-axis