cSQZ General
jacques - posted 21:59, Sunday 15 October 2023 (11365)
Beam profile of GR after SHG

NOTE : This measurement was performed before adding the EOM to the 1064 incident path to the polarization box, although it should not have a big consequence on the green light's profile (but may induce some mode mismatch from the incident 1064 to the SHG cavity). 

On Wed 10/04 I measured the beam profile of the GR generated from the SHG. The measurements were taken from the fraction of the GR which is recflected off the first DBS after the SHG.

Figures 1 and 2 show the beam profile with numerical fit for waist and M2, with and without the extra converging lens. The origin of the x scale is taken at the extremity of the csqz table, i.e. when the GR quits the table after being reflected from the DBS.

There is some ellipticity which is expected as the SHG crystal is birefringent.

 

 

Images attached to this report