GRAVITES General
epolini - posted 21:02, Friday 13 October 2023 - last comment - 18:37, Wednesday 18 October 2023(11364)
System realignment and TF, error and correction signal closed loop measurements

At some point today the system disaligned and the beam coupling in fiber was completely lost. I recovered the alignment of the system and I set the same optimized power values as before.

Then I took the OLTF of the whole system, shown in Fig. 1. Both curves are taken choosing a PI corner = 3kHz and LF gain limit = 90dB. I changed the gain (which has a broken knob so the value is not known a priori): a lower gain (blue curve) gives an UGF = 1700 Hz, an higher gain (orange curve) gives an UGF = 2500 Hz. The latter shows an oscillation around 20-30kHz that could be due to a too high gain or to a lack of coherence. 

In Fig.2 the OLTF acquired some days ago with similar settings PI corner = 1kHz (blue) and 3kHz (orange) and LF gain limit = 90dB. Here the UGF is higher wrt today: UGF(PI = 1kHz) = 8300 Hz,  UGF(PI = 3kHz) = 4900 Hz. The gain was the same.

Today I also acquired the error and correction signal spectra in closed loop, but unfortunately after I turned off and on again the laser. I checked that the system was in the same configuration as beforehand. 

In Fig. 3 the error signal (blue) and the correction signal (orange) closed loop for a PI corner = 3kHz are shown.

Then I acquired the TF for both PI corner of 3kHz (blue) and PI corner of 10kHz (orange) and they are shown in Fig. 5. Increasing the PI corner I get the weird behavior at high frequency as shown also before. The UGF in this case is 443 Hz for the blue curve and 949 Hz for the orange one.

In Fig.4, the corresponding error signal (blue) and the correction signal (orange) closed loop for a PI corner = 10kHz are shown (the lower resolution at low frequency is due to the fact that for some reason the SR785 did not save the last file of data).

The fact that the TF changes so much for different times of the measurements is weird and I am going to investigate it in the next days.

Images attached to this report
Comments related to this report
epolini - 18:37, Wednesday 18 October 2023 (11372)

After speaking with Eric, I realized that one possible issue of this measurement was where the input for the TF was taken. 

I was taking the output for the TF from the -B channel of the PID LB1005, same as the swept sine injection. I should indeed take it from the Error Monitor channel. 

So in summary I need to connect :

  • PID channel A to mixer output and to the channel B of SR785
  • PID channel -B to the trigger out of the SR785 (swept sine noise injection)
  • PID Error Monitor channel to the channel A of SR785.

In Fig.1 the OLTF of the system is shown for the following settings: PI corner = 3kHz, LF gain limit = 90dB and reasonable gain. The loop has an UGF = 16.8kHz, which makes more sense according to the correction and error signal closed loop I have been measuring.