Today I measured the dark noise of the new focus 1811 RF PD used in the unbalanced MZI. I connected the RF signal to the mixer together with the RF LO from the generator, then the output is filtered with the 1.9MHz filer.
I first measured on the oscilloscope the signal with the laser light on, to determine the calibration factor from Volt to radians: ASD[V] = A * ASD[rad], where A is the peak amplitude of the signal, which is A = 0.255 V/rad. Then I measured the spectrum of the output of the mixer with the SR785 without laser light on the PD, in Vrms/rtHz. The spectrum calbrated in rad/rtHz is shown in Fig. 1.