Reports until 17:38, Thursday 05 October 2023
epolini - posted 17:38, Thursday 05 October 2023 - last comment - 18:54, Wednesday 11 October 2023(11356)
How to measure TF with the SR785

[Eleonora, Wen]

We detail here the instructions to perform a TF measurement with the SR785, how to extract / save data on a USB stick or floppy and how to convert the file.78D to file.asc.

Display Options -› Format -> Dual

Freq -> Type -> Log

Display Setup -> Measurement Group -> Swept Sine

Display Setup -> Meausrement -> Freq Resp

Active display to change from the first to the second subscreen

First Display: Display Setup -> View -> Log Mag

Second Display: Display Setup -> View -> Phase or Unwrp. Phase

Select the span in the frequency menu (Start and Stop)

Select the resolution of the swept sine in the average menu

Select the sweped sine amplitude in Source -> Amplitude

Connect input in Channel one A

Connect output in Channel two A

Save:

Disk -> File name -> Alt -> type file name

Disk -> Display to Disk

Convert data with the software on the NW22 control room computer (from floppy or USB stick):

Open Command Prompt

C:\SRTRANS> srtrans.exe ./FolderName/*.78D

Comments related to this report
epolini - 18:54, Wednesday 11 October 2023 (11359)

It is possible to save the data in .TXT, with the following commands:

Output -> Hard Copy Button ->  ASCII Dump

Output -> File start Number -> select output number

It saves the file as SCRN0001.TXT.