[Eleonora, Wen]
We detail here the instructions to perform a TF measurement with the SR785, how to extract / save data on a USB stick or floppy and how to convert the file.78D to file.asc.
Display Options -› Format -> Dual
Freq -> Type -> Log
Display Setup -> Measurement Group -> Swept Sine
Display Setup -> Meausrement -> Freq Resp
Active display to change from the first to the second subscreen
First Display: Display Setup -> View -> Log Mag
Second Display: Display Setup -> View -> Phase or Unwrp. Phase
Select the span in the frequency menu (Start and Stop)
Select the resolution of the swept sine in the average menu
Select the sweped sine amplitude in Source -> Amplitude
Connect input in Channel one A
Connect output in Channel two A
Save:
Disk -> File name -> Alt -> type file name
Disk -> Display to Disk
Convert data with the software on the NW22 control room computer (from floppy or USB stick):
Open Command Prompt
C:\SRTRANS> srtrans.exe ./FolderName/*.78D
It is possible to save the data in .TXT, with the following commands:
Output -> Hard Copy Button -> ASCII Dump
Output -> File start Number -> select output number
It saves the file as SCRN0001.TXT.