GRAVITES General
epolini - posted 11:46, Wednesday 27 September 2023 (11347)
Termistor calibration check

Since the temperature shown by the termistor was not correct, I investigated the calibration of the termistor checking the read-out circuit with the help of Alvaro's thesis. The termistor is a TDK G1540/10k/F Negative Temperature Coefficient (NTC) and from its datasheet we can derive the relationship between the resistance R and the temperature T, as shown in Fig. 1. The circuit used to read out the temperature as a voltage is shown in Fig. 2. I verified that the resistors in this drawing correspond to the resistors in the integrated circuit and they do. The Vout is given by: Vout = (R2 + R3)/R2 * R1/(Rntc + R1) * Vdd - Vref * R3/R2, where R1 = 7.5 kOhm,  R2 = 1 kOhm, R3 = 6.8 kOhm, Vdd = 9 V, Vref = 5 V and Rntc = 0.006 * T**2 - 0.67*T + 23.408 from the fit in Fig.1. Vout as a function of the temperature is shown in Fig. 3 and I considered a restricted range (15°C - 40°C) wrt Alvaro's thesis (15°C - 55°C), to have a better linear fit: Vout = 0.6*T - 19.5. The calibration is consistent to what was used before and according to Eric a possible problem of this circuit could be that the resistor R1 is heated by the current passing through it, increasing the overall temperature. This would not explain the discrepancy we were reading yesterday, but it could be a reason why the Vout is higher than what we expect. 

I put back the termistor on the table and I taped it. I recorded the data with the cymac and in Fig. 4 the last hour of temperature data are shown on ndscope (checkign the X5:ISC-STATE_OBS4_INMON, sampled at 16Hz). The absolute value in counts can be converted in Volts and then in temperature with the following formulas:

Volts = counts * 40V / 2^16

T = (Vout + 19.512) / 0.6 ~ 25°C.

The temperature in the lab is around 20°C, meaning that the absolute value cannot be trusted. I will install some easier monitor system to check longterm temperature fluctuations. In the last hour the temperature changes are of the order of 0.12V which corresponds to a change in temperature of 0.2°C, which makes more sense wrt last time. Maybe the problem before was in the acquisition of data with the matlab script or some wrong connection in the circuit during the last data acquisition time.

Images attached to this report