Readout setup shifted from mirror B (power at readout 0.02mW) to mirror C (power at readout 4.5mW)
Readout PD (new focus 1811) spectrum at low power:
- Transmission at carrier as well as PDH sideband frequency (16.8MHz)
- ~100kHz comb type feature at carrier, repeats around sideband
Shifting to higher power can some what hide the features, but it still pushes noise above shot noise
Spectrum of PDH reflection PD (PDA10CS) when cavity is unlocked, no signal to PZT:
- ~100kHz lines around the 16.8MHz sideband
- If signal to EOM is turned off, the spectrum is flat
- If internal path of cavity is blocked, only the 16.8MHz sideband line shows up
- => Some frequency fluctuation caused by cavity?