Reports until 19:00, Friday 22 September 2023
GRAVITES General
epolini - posted 19:00, Friday 22 September 2023 (11340)
RIN measurement with SR785

As described in the last entry, the RIN cannot be measured with the cymac without adding an amplifier (i.e. SR560). 

In the meanwhile, we measured it with a SR785 signal analyzer, AC coupled and floating, with input gain range -32dBVpk.

The DC voltage read by the PD is 4.3V. 

The intensity noise floor is 2.6uVrms at few kHz: RIN = 20*log10(2.5e-6/4.3) = -125dBc/Hz.

The dark noise floor (of the Thorlabs PDA10CS) is 600nVrms : dark noise =  20*log10(600e-9/4.3) = -137dBc/Hz.

From the datasheet we derived the detector noise and the shot noise, considering that the PDA10CS was set to gain = 30dB.

Detector noise = NEP * responsivity * Gain = 1.25e-12 W/rtHz * 0.8 A/W * 4.75e4 V/A = 4.75e-8 V/rtHz

Shot noise = sqrt(2 * I * e) * Gain = sqrt(2 * 4.3 V / 4.75e4 V/A * 1.6e-19 A/Hz) * 4.75e4 V/A = 2.55e-7 V/rtHz

The limiting noise in this measurement is the shot noise and it has a factor 2 discrepancy wrt the dark floor noise (600nV /rtHz = 6e-7 V/rtHz).