As described in the last entry, the RIN cannot be measured with the cymac without adding an amplifier (i.e. SR560).
In the meanwhile, we measured it with a SR785 signal analyzer, AC coupled and floating, with input gain range -32dBVpk.
The DC voltage read by the PD is 4.3V.
The intensity noise floor is 2.6uVrms at few kHz: RIN = 20*log10(2.5e-6/4.3) = -125dBc/Hz.
The dark noise floor (of the Thorlabs PDA10CS) is 600nVrms : dark noise = 20*log10(600e-9/4.3) = -137dBc/Hz.
From the datasheet we derived the detector noise and the shot noise, considering that the PDA10CS was set to gain = 30dB.
Detector noise = NEP * responsivity * Gain = 1.25e-12 W/rtHz * 0.8 A/W * 4.75e4 V/A = 4.75e-8 V/rtHz
Shot noise = sqrt(2 * I * e) * Gain = sqrt(2 * 4.3 V / 4.75e4 V/A * 1.6e-19 A/Hz) * 4.75e4 V/A = 2.55e-7 V/rtHz
The limiting noise in this measurement is the shot noise and it has a factor 2 discrepancy wrt the dark floor noise (600nV /rtHz = 6e-7 V/rtHz).