Reports until 15:22, Friday 22 September 2023
GRAVITES General
epolini - posted 15:22, Friday 22 September 2023 (11338)
RIN measurement attempt with cymac

[Eleonora, Nikhil]

I tried to measure the RIN of the NPRO laser (19th Sept ET, 20th Sept UTC) and compare it with the dark noise, as shown in Fig. 1 for the following GPSs:

Fig. 1 shows the results for the RIN, i.e. the PSD normalized by the mean of the laser power data. We are limited by some electronics nosie.

Yesterday (21st September 2023), we decided to take other data to understand where the noise was coming from. First of all, we increased the PD gain from 20dB to 30dB, in order to read on the oscilloscope 4.6V instead of 1.6V, a further increase of the gain leads to PD saturation. 

In the following I list the GPSs times for different measurements (always using OBS4 channel on the cymac):

Fig. 2 shows the FFT spectra of these measurements. The wire is introducing extra noise. As expected, the cymac has a noise floor of 4uV/rtHz.

We acquired again the laser intensity noise and dark noise measurements:

Fig. 3  shows the FFT spectra of dark noise (blue), laser intensity noise (orange) and cymac noise (yellow). It is clear that the floor of this measurement is limited by the cymac + wire noise. 

We measured the RIN with the SR785 analyzer for the moment.

Images attached to this report