Reports until 13:49, Friday 22 September 2023
GRAVITES General
epolini - posted 13:49, Friday 22 September 2023 (11337)
Unbalanced MZI error signal acquired with the cymac

Yesterday (21st September 2023) I acquired data of the MZI's error signal in different configurations with the only channel available on the cymac (X5:ISC-STATE_OBS4_IN1_DQ).

Error signal with the loop OPEN taken after the mixer

In Fig. 1, the error signals for different GPSs are shown: at 19:28 UTC the peak is at 894Hz, at 19:32 UTC the peak is at 2827Hz, at 19:36 UTC the peak is at 4940Hz. I cannot acquire a spectrum higher than 16kHz with the cymac in this configuration, so I didn't wait the frequency difference to get higher. For each of these spectra I take 20 seconds of data. 

Afterwards, I tuned the RF generator's frequency to bring the error signal oscillations down to few hundreds Hz. Then I closed the loop.

Error signal with the loop CLOSED taken after the mixer 

Fig. 2 shows the two spectra (20 seconds of data each). On the spectrum analyzer, I checked the spectrum for higher frequencies. The DC is around -22dBm and then it goes down to -75dBm until stabilize arpund -100dBm after 200kHz. Which converting in V gives the same order of magnitude of what is shown in this plot.

Connecting the correction signal (the output of the PID) to the cymac was unlocking the loop. I left the PI corner at 10kHz.

Images attached to this report