epolini - posted 23:45, Tuesday 19 September 2023 (11330)
RIN measurement setup

Yesterday I set up a photodiode at the unused port of the PBS of the second power controller of the set up in order to record the amplitude fluctuations of the laser. I installed a steering mirror, a lens (focal length = 50mm) to focus the beam on the photodiode (Thorlabs PDA10CS) and two ND filters in front of the PD (ND10A and NE20A). After aligning the PD with the mirror, the power level recorded on the oscilloscope is 1.44V. Fig.1 shows the set up. 

Since only one channel of the cymac works properly, I disconnected the temperature sensor and I connected the PD to the cymac ADC (around 8pm ET). Nikhil changed the sampling frequency of the recorded data for the channel X5:ISC-STATE_OBS4_IN1_DQ, from 512Hz to 32768Hz. 

GPS of data:

 

Images attached to this report