GRAVITES General
epolini - posted 14:57, Tuesday 19 September 2023 (11328)
Unbalanced MZI visibility estimate and paddles installation

[Eleonora, Vicky]

Yesterday, I tried to measure the visibility of the unbalanced MZI. Since the output is coupled to a PD with two ports, one RF an one DC, it was not possible to measure it by just looking at the RF fringes (71MHz). This means that the DC component is filtered out from the RF signal and the RF fringes are centered around zero. In order to get a rough estimate of the visibility, I balanced the power in each arm and I measured the distinct power of each arm on the oscilloscope and the amplitude of the peak-to-peak of the RF fringes. Then we converted them into mA taking into account the different Transimpedance Gain for DC (10V/mA) and AC (40V/mA) according to the datasheet. Beforehand, I installed a polarization controller (paddles) l on the long arm of the MZI (see Fig. 1) in order to match the polarizations in the two arms by maximizing the fringes' amplitude. The peak-to-peak amplitude increased from 0.6V to 1V by adjusting the paddles' angles. 

DC voltage and current:

V_DC1= V_DC2 = 0.150V -> V_max = 0.3V

current = 0.3 V * mA / 10 V = 0.03 mA

AC voltage and current:

V_pp = 1 V 

current = 1 V *mA / 40 V = 0.025 mA

Considering that the visibility should be maximized when these two values are equal, we can conclude that an estimate of the visibility could be 0.025/0.03 * 100 = 83%.

Since, the fibers of the two arms should be very similar (customized from the same company) and then the beams are recombined in a fiber BS, if the powers are balanced, we expect the visibility to be good. Moreover, we shouldn't have mode matching / alignment issues as in free-space. Polarization changes can affect the visibility, but it can now be recovered by manually acting on the paddles. 

Images attached to this report