epolini - posted 19:40, Wednesday 13 September 2023 - last comment - 14:55, Friday 15 September 2023(11323)
Temperature monitor installed on the 200-meters-fiber spool

I moved from green lab the temperature monitor board, shown in Fig. 1 and I installed it next to the box containing the 200-meter-long fiber, shown in Fig. 2. I fixed the sensor to the plastic spool (Fig. 3) but I will also check it fixed on the bench. I am giving the +15V and the -15V with the Keysight E3630A Laboratory DC power supply, triple-output that was used to power the Rio fiber laser on the bench. I connected the sensor BNC output on the bigger box to the THERM1 BNC input of the smaller box and the TOUT1 before to the multimeter and then to the cymac. I am recording temperature data with the same channel used for the error signal (X5:ISC-STATE_OBS4_IN1_DQ).

The conversion Volt to Celsius for the temperature is (Volts+22.7)/0.601 = X deg C.

Images attached to this report
Comments related to this report
epolini - 18:12, Thursday 14 September 2023 (11324)

Fig. 1 shows the temperature drift over last night, while the sensor was placed on the plastic spool inside the box. 

The temperature changed by 0.065°C over more than 15 hours which corresponds to a 0.17% change in 15 hours and 0.01% change in 1 hour.

Images attached to this comment
epolini - 14:55, Friday 15 September 2023 (11325)

Yesterday I moved the thermistor to the bench next to the spool, as shown in Fig. 1. Data acquired with the cymac are shown i Fig. 2 for 15 hours (from 09/14 UTC 20:00:00). There are some oscillations, 2 per hour, i.e. 0.5mHz, as zoommed in in Fig. 3. 

Images attached to this comment