In the past days I worked on the locking of the fiber length stabilization lock. The main updates with respect to the last logbook entry #11312 are listed in the following.
1. RF generator frequency and VCO frequency synchronized
By mixing them, I reduced the offset and closed the loop with the New Focus PID controller. Then I reconnected the original loop elements.
2. Optimization of the power coupling in fiber
The coupling efficiency in the two fibers reduced a lot over time. I optimized it back and with Vicky we replaced the mount of the short-fiber port's post since it was not stable (magnetic base one). We optimized the power reaching the RF PD by tuning the power controller before the two steering mirrors in front of the first MZ BS. We maximized the coupling efficiency by looking at the DC OUT and then we tuned the HWP in order to have the higher power that does not distort the RF OUT of the PD. We are then combining this signal with the LO RF generator set at 7dbm.
3. Filters between the VCO and the amplifier
As already mentioned, the VCO is set to 71MHz with the PID controller's internal generator. In order to filter the frequencies higher than 80MHz, we installed a filter DC-190MHz (BLP-200+). This filter is followed by two power attenuators 10dB and 1dB.
4. Reduce the drift between the VCO and RF generator
The error signal (mixer's output after LP at 1.9 MHz) oscillation frequency drifts over time. From hundreds of Hz till hundreds of kHz. To be able to engage the loop, we need to reduce this frequency by tuning the RF generator frequency down to hundreds of Hz.
5. PID controller settings to lock
The best configuration to engage the loop is PI corner of 300 Hz and LF Gain Limit of 80 dB. The knob of the gain is broken, so it is set by hand every time. I put the long fiber inside a metallic box to isolate it from the environment. In Fig 1 the error signal in blue and the correction signal in yellow are shown. The spikes correspond to my tapping on the metal box. The DC correction is around 7V, which is the DC value needed to set the VCO to its correct RF frequency.
The loop stayed locked over night for the past two nights. I checked the following day the situation in terms of frequency drift of the error signal and reduced it by adjusting the RF generator frequency and lock again.
Night between 13th and 14th September
Between the afternoon (around 6pm) of 13th September and the afternoon (around 3pm) of 14th September, the loop was still locked and responded to excitations to the box on top of the fiber spool. I verified that the error signal changed by stopping the light going to one of the two MZI arms and it did. Then I unlocked it and I verified the frequency of the error signal and it was around few hundreds of kHz. I reduced it to few hundreds of Hz and locked again.
Night between 14th and 15th September
Between 3pm of 14th September and 11am of 15th September the frequency drifted by 5kHz and I put it back to few hundreds of Hz (around 200Hz).