hudsonl - posted 15:30, Tuesday 12 September 2023 (11315)
OPO Locking Thermal Effects
Masaya and I noticed today that the cavity OPO scan measured by the red pitaya during lock aquisition looked visibly different from the cavity scan displayed on the oscilloscope when scanning the cavity length. We determined that the difference was due to the different scanning speeds (about 0.5 Hz during lock acquisition vs about 16 Hz when scanning the OPO and viewing the scan on the scope). For very slow cavity scans, the OPO seems to have a substantial thermal self-locking effect, which can be seen in the fact that the cavity transmission increases linearly and then drops sharply as we scan through a resonance, instead of having a symmetric, lorentzian, lineshape. 

A similar effect seems to cause the cavity scan's higher-order-mode content to change significantly dpending on the scan frequency, which can make lock acquisition more challenging if the apparent high-order mode content is large. This through off the OPO locking script this morning, and we implemented several changes to fix this problem.
(1) Scan the cavity at 16 Hz during OPO lock acquisition
(2) Do not use the curvefit linewidth as tolerance metric when setting the piezo back near resonance; this number will be thrown off by thermal effects that cause the spectrum to look highly non-lorentzian.