A few weeks ago, I switched the SHG cavity lock from dithering a cavity mirror to dithering the laser's frequency. This substantially improved the SHG's stability. Since then, I've spent a good bit of time attempting to switch the OPO over to a feedback loop that dithers the laser's frequency as well, but have run into several problems. (1) For moderate dithering amplitudes, the SNR of the error signal demodulated from OPO's green transmission PD is quite low. In general, the OPO's green transmission power is low, so the PD signal is fairly noisy. (2) If I increase the dithering amplitude enough to get an error signal with a decent SNR, this modulation level causes huge (>50%) modulations in the SHG's transmitted green power. (3) There are various PD gain and bandwidth problems. I modified one of the custom PDs in the setup (PD 14) to have a 6pF feedback capacitor and a 500k feedback resistor. This the large trsnsimpedence impedance gives close to the desired amount of amplification (5 Mega-Ohms). but has noise that rises linearly with frequency above about 5kHz. Below this corner, the noise is 1uV/sqrt(Hz). On the other hand, the PD-100A has a transimpedence gain of 5 Mega Ohms and noise of about 15uV/sqrt(Hz) at 8kHz, which is similar to the custom PD. At this gain level (70dB), the PD-100A has a low bandwidth of only 6kHz. To take advantage of dithering the laser frequency, we would like to dither at closer to 50kHz. For now, I plan to switch the OPO back to dithering one of the cavity mirrors. This may not be the optimal cavity lock, but it seems to work well enough and will allow me to begin aligning the squeezer's output beam to the homodyne.