epolini - posted 16:52, Friday 18 August 2023 - last comment - 12:32, Friday 25 August 2023(11293)
LASTI seismic noise study: pt 2

[Eleonora, Rich, Regina]

Yesterday and today we continued the work started in entry #11285. 

We tried to use the two T240 instruments in LASTI (one next to the squeezer and one near the BSC). We centered the masses of both of them and read the signals (X MON, Y MON, Z MON) both with an oscilloscope and on the CDS system: on the oscilloscope we detected some peaks if someone jumped next to the trillium but the spectra on CDS were flat. After several trials, including interfacing the electronics with a new instrument ( a Streckeisen STS-2 Broadband Sensor which is working), we concluded that the T240 instruments are not working. Rich will send one of the instruments to repair.

Installation of the Streckeisen STS-2

We installed a Streckeisen STS-2 seismometer under the desk in LASTI near the BSC, as shown in Figure 1 and we connected it ot the STS-2 Interface Chassis in the rack. We covered the STS-2 with a box (see Figure 2). NB. Everytime we want to move the seismometer we HAVE TO lock the internal masses with a screwdriver, otherwise we can break it.

We centered the masses by using the botton on the small green box which is connected to the instrument. The signals read by the Streckeisen STS-2 are reported in Figure 3 (M1:HPI-BSC_STSINF_B_{X,Y,Z}_OUT).

Streckeisen STS-2 calibration with L4C attempt

In order to check its calibration we want to compare it with the L4Cs. We connected the corner #4 of the HAMX to two L4Cs, one H and one V placed on the ground. The signals read in V and H were different as shown in entry #11285, which is weird. So we started investigating: we replaced the vertical L4C with a new one and it was showing a signal significantly lower. We measured the internal resistance of all the L4Cs and they were correct (5.5 kOhm). The L4Cs of the other corners sees a reduced motions due to the spring attuation. We decided to fix the four corners of the HEPI to ground in order to use the L4Cs already installed. The signals right after fixing the HEPI are shown in Figure 4, with H4 and V4 placed on the ground. We noticed that V3 was not connected to the electronics box and we decided to connect back the H4 and V4 channels to the original L4Cs. The signals after this actions are shown in Figure 5 and they look correct.

Now, we need to devide the signals by the response of the instrument. 

 

Images attached to this report
Comments related to this report
epolini - 12:32, Friday 25 August 2023 (11298)

Unfortunately, the L4Cs signals don't look as here anymore. Corners 3 and 4 produce signals that are one order of magnitude lower than corners 1 and 2 of HAMX. Today I took a spectrum of the L4C signals and IPS signals of HAMX since they share some electronics, shown in Figure 1. Except for IPS H1 and V1, the other IPS signals seems reasonable and consistent with each other. 

 

Images attached to this comment