hudsonl - posted 18:47, Thursday 17 August 2023 - last comment - 12:06, Friday 18 August 2023(11290)
OPO Locked with Properly Coated M3. IR Output is Much More Stable
I replaced the M3 mirror on the OPO with one coated with the design specifications. With the new M3 installed, I re-aligned the OPO and tuned up the dither lock. At the moment, the OPO dither lock is implemented by dithering a cavity mirror, but I expect to change this to a laser frequency dither lock when a pair of clocked red pitayas arrives next week.

The attached files show the 00 cavity mode on a camera with a 532nm band-pass filter, the IR (purple trace) and green (yellow trace) mode matching from a cavity scan.

Note: Previously, this post had a comment about IR stability in the OPO cavity. I believe this is likely resolved, but this claim wasn't supported by the plot I had included. I believe what was measured there was just the PD's dark noise and the OPO was not on dual resonance.
Images attached to this report
Comments related to this report
hudsonl - 12:06, Friday 18 August 2023 (11291)
As expected, the OPO is now much more stable in IR than it was before. The attached image shows the OPO's green transmitted power (yellow trace) and IR transmitted CLF power (purple/pink trace) with a 500ms timescale. We no longer observe the large IR power oscillations we saw before. This time around, I was careful to check that the cavity is on dual resonance and the power level on the IR PD is well above its DC offset output voltage.