I replaced the M3 mirror on the OPO with one coated with the design specifications. With the new M3 installed, I re-aligned the OPO and tuned up the dither lock. At the moment, the OPO dither lock is implemented by dithering a cavity mirror, but I expect to change this to a laser frequency dither lock when a pair of clocked red pitayas arrives next week.
The attached files show the 00 cavity mode on a camera with a 532nm band-pass filter, the IR (purple trace) and green (yellow trace) mode matching from a cavity scan.
Note: Previously, this post had a comment about IR stability in the OPO cavity. I believe this is likely resolved, but this claim wasn't supported by the plot I had included. I believe what was measured there was just the PD's dark noise and the OPO was not on dual resonance.