hudsonl - posted 18:54, Wednesday 16 August 2023 (11289)
Data Cleaning to Find OPO Dual Resonance with New M3
I swapped the improperly coated M3 for a new, properly coated M3, which has an IR reflectivity of 89% and a green reflectivity of 96%, compared to the opposite specs on the previous M3. With the new M3, we have more green transmission through the SHG, but less IR transmission. Since the IR transmission was already low, this makes it more challenging to identify co-resonant locations automatically. Before, the reflected IR power from a cavity scan could be immediately curve fit to find resonance locations, but now it needs to be processed first.

I implemented a python function that detrends the reflected IR power data and applies a band pass filter. This makes the IR resonance locations much easier to identify. The attached plot shows an example of this.