Today, I re-measured the beam profile after mirror M4 and obtained a result similar to the one reported in entry #11277 (with a confidence level of 3%). I installed the telescope described in the same entry. I took the two lenses with f = 100 mm and f = 50 mm from the 10-foot-long bench currently located in the green lab. To achieve a beam with a waist around 400 μm, I positioned the lenses at z1 = 3 cm and z2 = 20 cm from M4, respectively, as shown in Figure 1 and Figure 2. I measured the beam profile at the output of the telescope, considering z = 0 as the position of the second lens, and the data is shown in Figure 3. The beam waist is approximately 400 μm and is located 23 cm from the second lens. I placed the AOM around the waist.
The next steps will be to install the electronics as described in the entry #11269 and the BS/mirror combination shown in #11270 (but we run out of BS). The final goal is to stabilize the 200-meter-long fibers just arrived from OZ optics.
At the end, to be sure not to clip the beam in the AOM (diameter aperture = 0.5 mm), I replaced this telescope with a f = 77.3 mm lens placed 17.5 cm after the last steering mirror. The waist diameter is around 200 um and located at 12 cm from the lens. I centered the AOM's crystal around the waist position, being careful that the size of the beam was smaller than its active aperture along the propagation.
The AOM is driven with an RF generator for the moment and the optimal parameters set are: frequency = 80 MHz, amplitude = 3 dBm. The amplifier installed after the RF generator is powered with a 24V DC power. The input of the amplifier is 3dBm and its output is 29bBm, meaning that it has a gain of 24 dB (higher that the 20dB expected). Its compression point is 29.5 dBm.
I alignmed the AOM in order to maximize the power on the 1st order diffraction beam:
P(before AOM) = 63 mW
P(after AOM) = 62 mW -> P(0th order) = 22 mW and P(1th order) = 40 mW.
The diffraction efficiency is around 67%.
The shape of the beams exiting the AOM is shown in Figure 1: the 1st order is on the right and it is very elliptical and it diverges very fast. In order to measure the parameters of the beam exiting the AOM, I placed a 150mm lens distant 17.5 cm from the AOM and then I did the inverse matching on JamMt. The data on the different axis measured with the beam profiler are shown in Figure 2 (the reference position is the lens position). The average fit gives w0 (radius) = 96um and z0 = 34cm. It turns out that the waist is positioned 10 cm BEFORE the center of the AOM.
I used this beam to design a telescope on JaMmt to collimate the beam back to 2mm diameter. After several iterations I decided that the best solution (beam not propagating too long in free space, 2 lenses, 0th and 1st order beams separated enough etc) was:
where the positions are referred to the center of the AOM (z0 -10 cm). The resulting beam has w0 = 1113 um (the fixed collimator expects a diameter of 2.4mm) and z0 = 3.77m.
I also thought that it might be better to couple both path in fiber and recombine them on a fiber coupler before being detected to avoid to design and install additional telescope and other complications. The reference fiber is 10cm-long.