Dithering the laser's PZT allows us to dither at a much higher frequency than dithering a PZT on the optical cavities does. The attached plots show the error functions from dithering the laser's PZT at 80kHz, a factor of 10 greater than the previous 8kHz dither on the SHG and OPO PZTs. This should allow us to increase the loop bandwidths by about a factor of 10 for additional audio-band noise suppression. The attached plots are from the SHG. The laser's PZT can be driven by up to 100V, but is currently being driven by the Red Pitaya's output directly at 1V with no additional amplification. Considering that we want the laser's frequency range to stay well inside the cavity linewidth for the transmission dither lock, the ideal voltage level to the laser is about 6V, so the 1V drive is in about the correct range. In the future, we could dither harder with a 6x amplifier between the Red Pitaya's output and the laser if need be.