Control loop for locking: gain knob on LB1005 = 3.5 (s), 4 (p)
Open loop transfer function measurement:
- s polarization: UGF 80kHz, PM 18deg
- p polarization: UGF 12kHz, PM 60deg
Finesse measurement in p:
- Done by driving PZT with a 10Vpp triangle wave at 5Hz and taking measurement of Lorentzian at transmission. PZT coefficient ~4MHz/V
- Linewidth = 155kHz, FSR = 65MHz ⇒ Finesse = 400
Finesse measurement in s:
- Using ringdown method: turned off laser and took cavity storage time data on transmission
- Storage time tau = 48.7us ⇒ Finesse = 10e3