Moved setup into red lab after doing some noise spectrum analysis at the desk with the setup using the OP232, f = 40.0mm lens, and the 1" retroreflector (pictured below). Noticed the low freq noise appears to be higher by a factor of ~10 with this setup compared to when the LED was positioned immediately beside the QPD (with no lens), this did not change after moving into the lab. Next steps are to mechanically constrain the components more, mostly the LED, and consider other configurations which could miniaturize the apparatus more.
Noise Budget Calculations from last week:
Calculations of noise in QPD circuit components assuming use of an AD8675 op amp and the following values:
| voltage noise density e_n (AD8975) | 2.8 nV/sqrt-Hz |
| current noise density i_n (AD8675) | 0.3 pA/sqrt-Hz |
| temperature T | 298 K |
| feedback resistor R_f | 2000 ohms |
| photocurrent I | 1 mA |
| Noise | Value (nV/sqrt-Hz) |
|---|---|
| Shot Noise | 80 |
| Johnson Noise | 5.7 |
| Amp Voltage Noise | 3.5/f + 2.8 |
| Amp Current Noise | 0.6 |
| Total Noise (at high freq, added in quadrature) | 80.3 |
Attached shows a rough schematic of how we want the QPD readout portion of the RetroSense head circuit to look. Some guiding principles: