Displaying report 1-1 of 1.
Reports until 17:26, Friday 23 June 2023
RetroSense General
msredden - posted 17:26, Friday 23 June 2023 - last comment - 17:46, Monday 26 June 2023(11235)
Sensitivity comparisons with varied configurations
Measured the sensitivity and relative intensity of the sensor under different configurations including with/without an f = 25.6mm lens, with/without a 1.5mm iris, and using an OP232 and VSLY5850 LED. 

Found that spot size on the QPD had a significant effect on the sensitivity. Setups with the lens for both LEDs and the iris for the VSLY5850 produced the best results. 
Non-image files attached to this report
Comments related to this report
msredden - 17:46, Monday 26 June 2023 (11237)
6/26/2023 update

Tested the OP232 with the new .5" f = 40.0mm lens and the PLA mount. Found a relative intensity of 1.17 (compare to the OP232 with the 1" f = 25.6mm lens which had a relative intensity of 1.19). 
Non-image files attached to this comment
Displaying report 1-1 of 1.