evanhall - posted 18:11, Wednesday 21 June 2023 - last comment - 09:35, Monday 17 July 2023(11233)
ADBC laser swap

Swadha, Evan

Previous laser: NP Photonics Rock (see MIT:11213)

Current laser: Lighwave/JDSU/Lumentum NPRO

The laser worked fine for the first week but starting a few days ago would only run for about an hour before going into a cycle of repeatedly shutting itself off every 30 seconds or so. There is no indication of thermal overload on the diagnostic fault pin, but nonetheless I suspect more heat sinking is needed. We clamped the head to a ~2" thick aluminum block that sits directly on the table. So far it has been running for 2 hours without issue, putting out around 800 mW.

Comments related to this report
evanhall - 14:45, Sunday 09 July 2023 (11254)

Myron fabricated a new mount, but the behavior eventually recurred when the laser was left on for a few hours. We tried another aluminum mount, but the behavior recurred with that too.

To test whether this is a problem with the controller or the head, Swadha swapped the controller to JDSU FQN4113040 (manufactured May 2013).

swadha - 09:35, Monday 17 July 2023 (11260)

Ran for ~2 days with the swapped controller (JDSU FQN4113040) without shutting off. It would seem the problem was with the controller.